3.4.    Blood and BTEX Operators

3.4.1.   Introduction

The configuration of the circulatory operators in GENTEX is shown in Fig. 3.1.1. Of the operators shown, all large vessel operators including input tubing, artery, arterioles, venules, vein and output tubing, are modeled with blood operators. The exchange units are modeled with BTEX operators. Parameters related to the circulatory operators are listed in Table 3.4.5. Note that the labels of the variables shown on the XSIM model graphs may be different from the full XSIM parameter name, which must be unique.

Blood flow through vascular and BTEX operators

The total blood flow and the large vessele hematocrit are set in the GENTEX model layout window ( Fig. 3.4.1). The total blood flow is delivered to the input tubing, artery, vein, and output tubing. The fraction of this flow that is delivered to each pathway is controlled by the number of paths and the flow heterogeneity. All the flow delivered to a path goes to the arteriole, BTEX unit, and venule.


Extracted pic [1]

Figure 3.4.1.    Main GENTEX window.

3.4.2.   Blood operators

Introduction

In GENTEX, a nonexchanging vessel is defined as a large vessel that does not exchange with extravascular regions. However, exchange between the RBC's and the plasma is allowed in these "nonexchanging" vessels modeled by blood operators. Each operator is characterized by two parameters: volume, V, and relative dispersion, RD. Other parameters governing exchange and chemical reactions in the blood are from the capillary-tissue unit. Fig. 3.4.2 shows a XSIM control window of such an operator.

The maximum value permitted for relative dispersion is 0.48, and the minimum value is 0. Note that the blood operator approximates the desired RD with axial diffusion coeffcients.

The volume should be the physical volume of that component of the circulation in ml g-1. If the volume of an operator is set to 0, it is effectively removed from the system.

Arterial operators

The vascular operators on the arterial side are the inlet tubing, artery, and arterioles. For each tracer or nontracer substance, there is a single inlet tube and artery. Separate operators are provided for these two components for use when the tissue content of the tracer and nontracer (residue) is being observed. The inlet tube can provide delay and dispersion of the injection before the marterial reaches the tissue that is included in the residue signal.

There is one arteriole ( Fig. 3.4.2) for each of the flow paths. The volume specified for this component is divided equally amongst the paths.


Extracted pic [3]

Figure 3.4.2.    The arteriol operator window.

Venous operators

The vascular operators on the venous side are the venules, vein, and outlet tubing. As with arterioles, there is one venule per path, and the volume is distributed equally. The comments made above regarding the inlet tubing and artery also apply to the vein and outlet tubing.

3.4.3.   BTEX operators

Introduction

The capillary-tissue units of each path are modeled with BTEX operators. A different type of BTEX operator is used for each tracer depending on the number of regions required to model the behavior of that tracer. Detailed information about the BTEX operators is contained in Bassingthwaighte, Chan and Wang (1992); information about the computer implementation of a typical three region model is available on-line (man btex30).

The generic parameters of a BTEX model are shown in the table below. The names used in GENTEX are prefixed with the tracer label (r: RBC tracer, v: vascular or plasma tracer, e: extracellular tracer, p: permeant tracer and nontracer substance, no prefix: all) and postfixed by the region label (cap: capillary, rbc: red blood cells, p: plasma, ec: endothelial cell, isf: interstitial fluid, and pc: parenchymal cell).

Name

Description

V

Volume of the region, ml s-1

rvel

RBC to plasma velocity ratio.

D

Axial diffusion constant in the region, cm2s-1

For all extravascular regions (ec, isf, and pc) the volume is a volume of distribution rather than a physical volume. Note that erythrocyte mean transit time through the capillary system is less than plasma transit time, since the velocity of red blood cells is faster than that of plasma. The ratio of RBC to plasma velocity, rvel, is reflected in a difference between large vessel hematocrit (Hct) and capillary hematocrit (Hctcap). In this model, the total blood flow (Fb), capillary volume (Vcap), large vessel hematocrit (Hct), and RBC to plasma velocity ratio (rvel) are user-defined parameters. The intracapillary RBC flow, Frbc, and plasma flow, Fp, are calculated and identical to the large vessel flows, i.e., Frbc = Hct Fb and Fp = (1 - Hct) Fb. The volumes of RBC's and plasma (Vrbc and Vp) are calculated as Vrbc = Hctcap Vcap and Vp = (1 - HctcapVcap where Hctcap is the capillary hematocrit and is derived as follows: because


Extracted pic [13] ,




Extracted pic [14] ,

    


Extracted pic [15] ,


Extracted pic [16] ,

    and


Extracted pic [17]

so


Extracted pic [23] .

Because the species can move to all regions, it has an apparent volume of distribution (V') in all regions. In the RBC and plasma region, the volumes of distribution are


Extracted pic [18]

and


Extracted pic [19]

where Wrbc and Wp are the water fractions of the RBC's and the plasma, respectively. Note that W's are the contants that convert the physical volumes to volumes of distribution. They are referred as water fractions in GENTEX for convenience. See Fig. 3.4.4 for details of these model parameters.

RBC reference tracer

The RBC reference tracer is modeled with a one region BTEX model. The tracer is constrained to the RBC region.

Vascular reference tracer

The vascular tracer is modeled with a two region BTEX model ( Fig. 3.4.3). While the vascular tracer is nominally constrained to the plasma region, the second region is used to account for any binding of the tracer to the endothelial cell. An example of this effect is binding of albumin to an albumin-fatty acid receptor on the endothelial cell membrane. If appropriate, this second region can also be used to account for slow leakage of the vascular tracer into the ISF over long times. See Table 3.4.6 for details of these model parameters.


Extracted pic [2]


Figure 3.4.3.    BTEX model parameters for vascular tracer.

Extracellular tracer

The extracellular tracer also has access to two regions, plasma and ISF ( Fig. 3.4.4). For this tracer, however, the ISF is partitioned into two parts. This partitioning can be used to account for radial diffusion in the ISF or physical division of the ISF into two distinct regions (e.g., T-tubules in skeletal muscle). Apportionment of the total ISF volume between these two partitions is controlled by the parameter fVisf2 and the diffusion of tracer between them by the parameter PSisf. If partitioning of the ISF is not desired, set both fVisf2 and PSisf to 0. See Table 3.4.7 for details of these model parameters.


Extracted pic [12]

Figure 3.4.4.    BTEX model parameters for extracellular tracer.

Permeant species

GENTEX allows both tracer and nontracer mother substances of multiple permeant species (up to 5). GENTEX models linear and nonlinear processes of transmembrane transport and chemical reactions. Model parameters can be set up by clicking on the PS buttons for transmembrane transport, on the G buttons for consumption, and on the Bind buttons for binding ( Fig. 3.4.5). One can also access all GENTEX functionalities via the pull-down menu of Parameter in XSIM main window (See an example in Fig. 3.4.6). See Table 3.4.8 and Table 3.4.9 for details of the model parameters.


Extracted pic [4]

Figure 3.4.5.    Permeant tracer BTEX unit window.


Extracted pic [21]

Figure 3.4.6.    XSIM pull-down menu

3.4.4.   Transport and chemical reactions of permeant species in BTEX units

Transporters

There are transports at the membrane of RBC's, the luminal and abluminal membranes of endothelial cells, and the membrane of the parenchymal cells. Fig. 3.4.7 shows the control window for the transport through the luminal membrane of endothelial cells (click on the PSecl button in the BTEX Unit window for any permeant species).


Extracted pic [6]

Figure 3.4.7.    Control window for transmembrane transporter.

For each species, users can select one of the four transport types (T_type) ( Table 3.4.1). For Linear transport the constant PS values can be set in the main control window. For Facilitated and Nonlinear transport, further setup is required (clicking on the buttons at the bottom of the window).   

Table 3.4.1.:    Transporter type.
Labels Value Descriptionn


None

No transport


Linear

Linear, asymmetric

T-type

Facilitated (specific)

Species-specific, carrier-mediated, no receptor


Facilitated (non-specific)

Competitive, carrier-mediated, no receptor


Nonlinear

Competitive, carrier-mediated, receptor

Both Facilitated and Nonlinear transporters in GENTEX are carrier-mediated, i.e., substrate binds to the transporter at one side of the membrane, undergoes conformational change, and then is released at the other side of the membrane. The binding is assumed to quasi-equilibrium, i.e., much faster than the conformational change. The conformational change is assumed to be first-order with rate constants PT for free transporters and PTS for bound transporters. The main difference in Facilitated and Nonlinear transport is that the latter is more general and also has surface receptors.

The Facilitated transporter ( Table 3.4.8) in GENTEX can be species-specific or shared by different species (competitive). Its conformational change is always symmetric. It has zero transporter capacitance (no receptor). If two species are set up to use species-specific transporter, each species has its own set of transporter parameters. In the competitive transport case, one needs to choose Facilitated (specific) of T-type for the two or more species sharing the same transporter. The total transporter concentration (e.g., Ttot in Fig. 3.4.9 and Table 3.4.10) of the first species sharing the transporter is used in the model, i.e., only the first and one Ttot is in effect. Other transporter parameters can be species specific.


Extracted pic [5]

Figure 3.4.8.    Control window for the facilitated transporter.

A Nonlinear transporter ( Fig. 3.4.9) in GENTEX can be shared by multiple species. The conformational change can be asymmetric. The carrier also serves as a receptor. The parameters that are used by different types of transporters are listed in Table 3.4.10. The method for setting up competitive, facilitated transporter is applicable here.


Extracted pic [9]

Figure 3.4.9.    Control window for general nonlinear transporter (competitive, and with receptor)

Note that the effective PS values are calculated by the model. These values are averaged over all the axial segments and are shown in the control windows during a simulation run. 

Consumptions

For every region, there is a control window ( Fig. 3.4.10) for species consumption and transformation. Consumption type can be set separately for each species ( Table 3.4.2).


Extracted pic [8]

Figure 3.4.10.    Control window for consumption

Table 3.4.2.:    Consumption type

Label

Value

Description


None

No consumption


Linear

First-order reaction


Michaelis-Menten

Michaelis-Menten (M-M) kinetic type

G-type

Enzyme

Enzyme binding and reaction


M-M & Enzyme

Simultaneous reactions in M-M and enzyme kinetics


M-M & M-M

Simultaneous reactions in different M-M and M-M kinetics.

The parameters for these chemical reactions can be set up by choosing the G-typefirst, then provide correspoding parameter values. For linear and Michaelis-Menten reactions, the transformation of chemical species can be set up by using fractional rate in the middle part of the consumption control window. For example, a f23 of 0.4 means 40% of species #2 that is consumed will be transformed to species #3, and f22 is the sequestration fraction. GENTEX allows flexible configuration of reaction cascade but it is the user's responsibility to assure the mass balance.

Enzyme binding and reactions can be set up at the bottom of the consumption control window. The enzyme is shared if multiple species are thosen to use this reaction type.

GENTEX also allow multiple metabolic pathways of the same species by choosing G-type of M-M & Enzyme or M-M & M-M.

Table 3.4.12 and Table 3.4.13 provide the complete lists of consumption parameters for species #1 and #2 in parenchymal cells. Consumption parameters in other regions for other species are likewise.

Binding

For every region, one can use up to three binding sites for each species: B1, B2 and B3. B1 and B2 are fast binding sites (quasi-equilibrium), and B3 is a slow on-and-off binding site. For every region, there is a control window ( Fig. 3.4.11) for setting up the binding of all the species. Binding type can be set separately for each species ( Table 3.4.3). The full lists of parameters for species #1 and #2 in the ISF space are provided in Table 3.4.14 and Table 3.4.15. Binding parameters for other regions and species are likewise. Note that mobil binding sites that move with the flow are allowed in the RBC and plasma regions.

    


Extracted pic [7]

Figure 3.4.11.    Control window for binding.

  

Table 3.4.3.:    Binding type

Label

Value

Description

B1_type

None

No binding.

or

Fast/non-competi/immob

Fast, species-specific, stationary

B2_type

Fast/competi/immob

Fast, competitive, stationary


Fast/non-competi/mobile

Fast, species-specific, moving with flow


Fast/competi/mobile

Fast, compettive, moving with flow

B3_type

None

No binding.


Slow

Slow on-and-off, competitive, stationary

Bireactant reaction

In addition to the consumption and binding in every region, GENTEX allows bireactant reactions in one user-specified region by using the Bireactant reactions window ( Fig. 3.4.12; in the Main XSIM Window choose menu Parameters -> Consumptions -> Bireactant reactions). Users can select the region and set up individual bireactant reactions (up to 4),


Extracted pic [20]


where S's are the species, and "a" and "b" are fractional coefficients. Kf and Kr are the forward and reverse rate constants, s-1. Note that this feature applies to nontracer kinetics only, and Si, Sj and Sk must be different chemical species and tracer.


Extracted pic [10]

Figure 3.4.12.    Control window for bi-reactions

Species synthesis and time-varying parameters

GENTEX allows point or line sources to simulate species synthesis. From the XSIM menu (Parameters -> Species synthesis/sources in BTEX unit), one can get the control window ( Fig. 3.4.13). Users can have two sources.


Extracted pic [11]

Figure 3.4.13.    Control window for species synthesis.

Users also need select two source functions for the two sources (two buttons at the bottom). For the Additive Source, the function values represent the accumulative changes over time, i.e., the area under the souce function curves are the total changes. For other sources, the function values represent the absolute concentrations over time.

Time-varying parameters

GENTEX allows allows time-varying model parameters. From the XSIM menu (Parameters -> Time-varying Paremeters in BTEX unit), one can get the control window (). The function values represent the fractional change of the parameter over time. Users can have two time-varying parameters simultaneously.


Extracted pic [22]

Figure 3.4.14.   Control window for time-varying parameters.

3.4.5.   Model switches

There are three switches in the GENTEX model layout window: (1) initial condition (I.C.) switch, (2) switch for computing tracer or nontracer or both, and (3) ODE solver switch.

For the ODE solver, GENTEX provides four methods: (1) a fifth-order Taylor method for linear problems; (2) LSODES, a variant of back differentiation formulae (BDF) for nonlinear stiff and sparse problems (Authors: Alan c. Hindmarsh and Andrew h. Sherman, 1987); (3) RADAU, a variant of implicit Runge-Kutta method for nonlinear stiff problems (Authors: E. Hairer and G. Wanner, 1996); and (4) an adaptive method trying to use the best-suited solver (default solver). Note that the adaptive method is safe but not necessarily the fastest.

3.4.6.   Vascular and BTEX operator errors and messages

[TO BE DEVELOPED]

3.4.7.   Reference tables for GENTEX parameters 

Table 3.4.4.:    Common parameters for GENTEX

XSIM Label

Description

Unit

Typical Value

XSIM Name

Fb

Blood flow

ml min-1g-1

1.0


Hct

Large vessel hematocrit

unitless

0.0~1.0

hctLV

Wrbc

Water fraction of RBC

unitless

0.65


Wp

Water fraction of plasma

unitless

0.94


Vcap

Capillary volume

ml g-1

0.03


rvel

RBC to plasma velocity ratio

unitless

1.0~3.0


Nseg

Number of axial segments

unitless

1~60

Nseg_int

capillary length

Capillary length

cm

0.1

clngth

     

Table 3.4.5.:    Parameters for large vessel operator

XSIM Label

Description

Unit

Typical Value

XSIM Name

Vtube_i

Volume of input tubing

ml g-1



RDtube_i

Relative dispersion of input tubing

unitless



Vart

Volume of arteries

ml g-1



RDart

Relative dispersion of arteries

unitless



Vartl

Volume of arteriols

ml g-1



RDartl

Relative dispersion of arteriols

unitless



Vvenl

Volume of venules

ml g-1



RDvenl

Relative dispersion of venules

unitless



Vven

Volume of veins

ml g-1



RDven

Relative dispersion of veins

unitless



Vtube_o

Volume of output tubing

ml g-1



RDtube_o

Relative dispersion of output tubing

unitless



    

Table 3.4.6.:    Parameters for vascular reference tracer in BTEX units

XSIM Label

Description

Unit

Typical Value

XSIM Name

Dp

Axial diffusion constant in plasma

cm2sec-1


v_Dp

PSg

PS for capillary-ISF exchange

ml min-1g-1

0

v_PSg

V'isf

ISF virtual volume

ml g-1

0.15

v_Visfp

C @ t=0

Initial concentration

arbitary


v_Czero

 

Table 3.4.7.:    Parameters for excellular reference tracer in BTEX units

XSIM Label

Description

Unit

Typical Value

XSIM Name

Dp

Axial diffusion constant in plasma

cm2sec-1


e_Dp

PSg

PS for capillary-ISF exchange

ml min-1g-1


e_PSg

V'isf

ISF virtual volume

ml g-1

0.15

e_Visfp

fVisf2

Fraction of ISF volume in ISF2

unitless


e_fVisf2

Disf

Axial diffusion constant in ISF

cm2sec-1


e_Disf

PSisf

PS for ISF1-ISF2 exchange

ml min-1g-1


e_PSisf

C @ t=0

Initial concentration, arbitrary units

arbitary


e_Czero

    

Table 3.4.8.:    Parameters for permeant tracer #1 in BTEX units

XSIM Label

Description

Unit

Value

XSIM Name

Drbc

Axial diffusion constant in RBC's

cm2sec-1


p_Drbc

Dp

Axial diffusion constant in plasma

cm2sec-1


p_Dp

V'ec

Endothelial cell virtual volume

ml g-1

0.01

p_Vecp

Dec

Axial diffusion constant in endothelial cell

cm2sec-1


p_Dec

PSg1

Linear PS from capillary to ISF

ml min-1g-1

1

p_PSg1

PSg2

Linear PS from ISF to capillary

ml min-1g-1

1

p_PSg2

V'isf

ISF virtual volume

ml g-1

0.15

p_Visfp

Disf

Axial diffusion constant in ISF

cm2 sec-1


p_Disf

PSisf1

Linear PS from ISF2 to ISF1

ml min-1g-1


p_PSisf1

PSisf2

Linear PS from ISF1 to ISF2

ml min-1g-1


p_PSisf2

fVisf2

Fraction of ISF volume in ISF2

unitless


p_fVisf2

V'pc

Parenchymal cell virtual volume

ml g-1

0.6

p_Vpcp

Dpc

Axial diffusion constant in parenchymal cell

cm2sec-1


p_Dpc

     

Table 3.4.9.:    Parameters for permeant tracer #2 in BTEX units

XSIM Label

Description

Unit

Value

XSIM Name

Drbc

Axial diffusion constant in RBC's

cm2sec-1


p_Drbc_2

Dp

Axial diffusion constant in plasma

cm2sec-1


p_Dp_2

V'ec

Endothelial cell virtual volume

ml g-1

0.01

p_Vecp_2

Dec

Axial diffusion constant in endothelial cell

cm2sec-1


p_Dec_2

PSg1

Linear PS from capillary to ISF

ml min-1g-1

1

p_PSg1_2

PSg2

Linear PS from ISF to capillary

ml min-1g-1

1

p_PSg2_2

V'isf

ISF virtual volume

ml g-1

0.15

p_Visfp_2

Disf

Axial diffusion constant in ISF

cm2sec-1


p_Disf_2

PSisf1

Linear PS from ISF2 to ISF1

ml min-1g-1


p_PSisf1_2

PSisf2

Linear PS from ISF1 to ISF2

ml min-1g-1


p_PSisf2_2

fVisf2

Fraction of ISF volume in ISF2

unitless


p_fVisf2_2

V'pc

Parenchymal cell virtual volume

ml g-1

0.6

p_Vpcp_2

Dpc

Axial diffusion constant in parenchymal cell

cm2sec-1


p_Dpc_2 

 

Table 3.4.10.:    Transporter parameters for species #1.

XSIM Label

Type

Description

XSIM Name

Lin PSecl1

Linear

Linear PS from plasma to endothelial cells, ml min-1g-1

p_PSecl1

Lin PSecl2

Linear

Linear PS from endothelial cells to plasma, ml min-1g-1

p_PSecl2

Ttot

Facilitated

Total transporter concentration in the luminal side of endothelial membrane, mmol g-1

p_Ttotecl

Kd

Facilitated

Equilibrium dissociation constant @ side 1, Molar

p_Tkd1ecl

Kd

Facilitated

Equilibrium dissociation constant @ side 2, Molar

p_Tkd2ecl

PTS

Facilitated

Rate of linear conformational change of bound transporter, s-1

p_P11ecl

PT

Facilitated

Rate of linear conformational change of free transporter, s-1

p_P01ecl

Ttot

Nonlinear

Total transporter concentration, mmol g-1

p_Ttotecl

Kd @ reg1

Nonlinear

Equilibrium dissociation constant @ side 1, Molar

p_Tkd1ecl

Kd @ reg2

Nonlinear

Equilibrium dissociation constant @ side 2, Molar

p_Tkd2ecl

PTS+

Nonlinear

Rate of conformational change of bound transporter, s-1

p_P11ecl

PTS-

Nonlinear

Rate of conformational change of bound transporter, s-1

p_P12ecl

PT+

Nonlinear

Rate of conformational change of free transporter, s-1

p_P01ecl

PT-

Nonlinear

Rate of conformational change of free transporter, s-1

p_P02ecl

 

Table 3.4.11.:   Transporter parameters for species #2.

XSIM Label

Type

Description

XSIM Name

Lin PSecl1

Linear

Linear PS from plasma to endothelial cells, ml min-1g-1

p_PSecl1_2

Lin PSecl2

Linear

Linear PS from endothelial cells to plasma, ml min-1g-1

p_PSecl2_2

Ttot

Facilitated

Total transporter concentration in the luminal side of endothelial membrane, mmol g-1

p_Ttotecl_2

Kd

Facilitated

Equilibrium dissociation constant @ side 1, Molar

p_Tkd1ecl_2

Kd

Facilitated

Equilibrium dissociation constant @ side 2, Molar

p_Tkd2ecl_2

PTS

Facilitated

Rate of linear conformational change of bound transporter, s-1

p_P11ecl_2

PT

Facilitated

Rate of linear conformational change of free transporter, s-1

p_P01ecl_2

Ttot

Nonlinear

Total transporter concentration, mmol g-1

p_Ttotecl_2

Kd @ reg1

Nonlinear

Equilibrium dissociation constant @ side 1, Molar

p_Tkd1ecl_2

Kd @ reg2

Nonlinear

Equilibrium dissociation constant @ side 2, Molar

p_Tkd2ecl_2

PTS+

Nonlinear

Rate of conformational change of bound transporter, s-1

p_P11ecl_2

PTS-

Nonlinear

Rate of conformational change of bound transporter, s-1

p_P12ecl_2

PT+

Nonlinear

Rate of conformational change of free transporter, s-1

p_P01ecl_2

PT-

Nonlinear

Rate of conformational change of free transporter, s-1

p_P02ecl_2

 

Table 3.4.12.:    Parameters for consumption of species #1 in parenchymal cells.
XSIM Label Type Description XSIM Name

Linear G

Linear

Linear consumption rate, ml min-1g-1

p_Gpc

M-M Gmax

Michaelis-Menten

Maximal rate in M-M reaction, mmol g-1

p_Gmaxpc

M-M Km

Michaelis-Menten

M-M constant, Molar

p_Gkmpc

f11

Linear or M-M

Fraction of species #1 sequestered

p_R22pc

f12

Linear or M-M

Fraction of species #1 transformed to species #2

p_R21pc

f13

Linear or M-M

Fraction of species #1 transformed to species #3

p_R23pc

f14

Linear or M-M

Fraction of species #1 transformed to species #4

p_R24pc

f15

Linear or M-M

Fraction of species #1 transformed to species #5

p_R25pc

Etot

Enzyme

Total enzyme concentration, mmol g-1

p_Etotpc

k+

Enzyme

Association constant, Molar-1 s-1

p_Ek1pc

k-

Enzyme

Dissociation constant, s-1

p_Ek_1pc

kf

Enzyme

Rate constant of forward reaction, s-1

p_Ekfpc

kr

Enzyme

Rate constant of reverse reaction, s-1

p_Ekrpc

 

Table 3.4.13.:    Parameters for consumption of species #2 in parenchymal cells.
XSIM Label Type Description XSIM Name

Linear G

Linear

Linear consumption rate, ml min-1g-1

p_Gpc_2

M-M Gmax

Michaelis-Menten

Maximal rate in M-M reaction, mmol g-1

p_Gmaxpc_2

M-M Km

Michaelis-Menten

M-M constant, Molar

p_Gkmpc_2

f21

Linear or M-M

Fraction of species #2 transformed to species #3

p_R21pc_2

f22

Linear or M-M

Fraction of species #2 sequestered

p_R22pc_2

f23

Linear or M-M

Fraction of species #2 transformed to species #3

p_R23pc_2

f24

Linear or M-M

Fraction of species #2 transformed to species #4

p_R24pc_2

f25

Linear or M-M

Fraction of species #2 transformed to species #5

p_R25pc_2

Etot

Enzyme

Total enzyme concentration, mmol g-1

p_Etotpc_2

k+

Enzyme

Association constant, Molar-1 s-1

p_Ek1pc_2

k-

Enzyme

Dissociation constant, s-1

p_Ek_1pc_2

kf

Enzyme

Rate constant of forward reaction, s-1

p_Ekfpc_2

kr

Enzyme

Rate constant of reverse reaction, s-1

p_Ekrpc_2

 

Table 3.4.14.:    Parameters for binding of species #1 in the ISF region.
XSIM Label Type Description XSIM Name

B1_tot

Fast

Concentration of binding site B1, mmol g-1

p_B1totisf

B1_Kd

Fast

Equilibrium dissociation constant for B1, Molar

p_B1kdisf

B2_tot

Fast

Concentration of binding site B2, mmol g-1

p_B2totisf

B2_Kd

Fast

Equilibrium dissociation constant for B2, Molar

p_B2kdisf

B3_tot

Slow

Concentration of binding site B3, mmol g-1

p_B3totisf

B3_Kd

Slow

Equilibrium dissociation constant for B3, Molar

p_B3kdisf

B3_K+

Slow

On-rate for B3, Molar-1 s-1

p_B3k1isf

  

Table 3.4.15.:    Parameters for binding of species #2 in the ISF region.
XSIM Label Type Description XSIM Name

B1_tot

Fast

Concentration of binding site B1, mmol g-1

p_B1totisf_2

B1_Kd

Fast

Equilibrium dissociation constant for B1, Molar

p_B1kdisf_2

B2_tot

Fast

Concentration of binding site B2, mmol g-1

p_B2totisf_2

B2_Kd

Fast

Equilibrium dissociation constant for B2, Molar

p_B2kdisf_2

B3_tot

Slow

Concentration of binding site B3, mmol g-1

p_B3totisf_2

B3_Kd

Slow

Equilibrium dissociation constant for B3, Molar

p_B3kdisf_2

B3_K+

Slow

On-rate for B3, Molar-1 s-1

p_B3k1isf_2


Copyright 1998-2000, National Simulation Resource, University of Washington, Author: Zheng Li and Rick King <zhengli@bioeng.washington.edu> Last modified: 26 Nov 2000

Previous Next Contents Sorry, No Index Feedback